| Accuracy: ±2°C temperature corrected |
| Repeatability: 0.8°C |
| Response time: 95% of step change in 3 min. |
| Sample flow: >1 LPM |
| Temperature Sensor | Range: -40°C to +70°C |
| Accuracy: ±2°C |
| Pressure Sensor (optional) | 2 to 300 PSIA ±1%, 316L stainless steel wetted parts, higher pressures can be ordered optionally |
| Flow meter | Body: Styrene Acrylonitrile, Max Pressure: 50 psi (3.4 bar) [ this pressure limitation is for the exhaust ] |
| Inlet Metering Valve | Max pressure: 300 psi (20 bar) [ this pressure limitation is for the inlet, and can be utilized only if outlet is unrestricted] |
| Ergonomics | Easy to carry or use on desktop, lightweight 2 Lbs |
| Easy to read graphic LCD display with user controlled backlight brightness (LCD is readable without the backlight) |
| Snap dome tactile membrane switches with dynamically assigned legends |
| Audible feedback upon button press as well as for warnings and alarms |
| Auto scaling graph/histogram on moisture measurement display for stability verification |
| Easy to use with verbose built-in context sensitive Help screens |
| Battery Saver, with user selectable Auto-power-down times, can also be disabled |
| Built-in pressure measurement and pressure corrected moisture readings, simultaneously displayed with the uncorrected reading for greater ease of understanding |
| Verbose Status bar with indications, warnings and audible errors |
| User selectable locking with or without a password. |
| Built-in battery backed clock calendar |
| Electrical | External AC Universal Power adapter (100-240VAC 50-60Hz) with USB connector |
| Rechargeable Li-Ion battery, recharges through the USB connector from the PC or from the supplied universal power adapter |
| Sampling system and sensor are isolated from electrical interface and supply grounds |
| Measurement resolution of moisture sensor: 1pF resulting in less than 0.01 ̊C |
| External power and I/O: USB mini B female connector – IP67 / IP68 with Sealing Cap |
| Alarm Relay Option – 3A, 250VAC30VDC contacts |
| AV alarm >80 dB @ 3′, 3 jumbo Red LEDs 5100 mcd each, complies with NFPA99, 1996 edition |
| Mechanical | Enclosure |
| ABS (UL94-HB), with EDPM gasket to meet IP65, NEMA-4X |
| Dimensions H: 5.71″(145mm) W: 4.13″(105mm) D: 2.56″(65mm) |
| Weight Total: 2 Lbs (~1 Kg) |
| 1/4″ NPT Female Swivel inlet, allowing connection with standard quick connect |
| Aluminum and Viton wetted parts, small surface area sampling chamber for fast response time. |
| Pressure: Inlet 300 psi (20 bar) Outlet 50 psi (3.4 bar) |
| Temperature Range | dewPatrol Analyzer: -20°C to +60°C |
| Electronics: -40°C to +85°C |
| Moisture Sensor: -20°C to +60°C |
| Pressure Sensor: -20°C to +85°C |
| LCD operating: -20°C to +70°C storage: -30°C to +80°C |
| Enclosure: -40°C to +60°C |
| Miscellaneous Features | NIST traceable calibrations |
| Units of measure: °C & °F dewpoint, ppmV, ppmW, µB H2O vapor pressure, grams of H2O / m3, |
| Lbs H2O /106 standard cubic feet in Natural Gas. (All units auto-scaled) |
| Digital & Analog I/O: USB, 4/20mA with HART® |
| Data logging – single point, multipoint, time stamping, each sample records: dewpoint, temperature and pressure, ( memory > 4,000 samples) |
| Virtual Analyzer PC software for training and evaluation, allows the user to experience the exact interface on their own PC with voice explanations |