Model 2500
Model 9500
Automated “Two-Pressure” Humidity Generation System
The Model 9500 Humidity Generator is a system capable of producing known humidity values using the fundamental, NIST proven, “two-pressure” principle. The 9500 is capable of continuously supplying an accurately known relative humidity, dew point, frost point, parts per million, or other calculated values for instrument calibration and evaluation as well as precision environmental testing. This system will automatically generate manually entered humidity and temperature set points as well as user created multipoint humidity and temperature profiles. All desired humidity’s, temperatures, test pressures, flow rates, and time intervals may be programmed. Visual indications of system status are displayed in real time on the computer monitor.
FEATURES
- 0.17% * R + 0.016 RH Uncertainty 5,6
- Traceable to SI 8
- High Flow Capability of 10 to 100 L/min
- Based on NIST Proven "Two-Pressure" Principle
- Generate: RH, DP, FP, PPM, Multipoint Profiles
- Computerized Internal Transducer Calibration
- Computes System Uncertainties in Real Time
- Automatically Applies Enhancement Factors
- 9500 ControLog® Embedded Automation Software
- HumiCalc® with Uncertainty Mathematical Engine
Relative Humidity Range (Test Chamber @ 14.7 psiA) | 5 to 98 %RH |
---|---|
Dew Point Temperature Range (Test Chamber @ 14.7 psiA) | –35 to 70 °C |
Frost Point Temperature Range (Test Chamber @ 14.7 psiA) | –32 to 0 °C |
Bath Temperature Range: 1 | 0 to 72 °C |
Bath Temperature Control Stability: 2 | 0.002 °C |
Bath Temperature Heating Rate: from 0 to 72 °C | 0.5 °C per Minute (average) |
Bath Temperature Cooling Rate: from 72 to 0 °C | 0.5 °C per Minute (average) |
Chamber Temperature Uniformity: 3 | 0.008 °C |
Temperature Specification: 0 to 72 °C 6 | ±0.015 °C |
Gas Type: | Air or Nitrogen |
Gas Pressure Rating: (MAWP) | 350 psiG |
Gas Flow Rate Range: | 10 to 100 L/min |
Gas Flow Rate Specification: | ±2% of full Scale |
Supply Pressure Specification: | ±1.25 psiG |
Saturation Pressure - Low Range: | Ambient to 45 psiA |
Saturation Pressure - High Range: | 45 to 325 psiA |
Low Range Pressure Specification: | ±0.0045 psiA |
High Range Pressure Specification: | ±0.03 psiA |
Test Chamber Pressure Range: | Ambient to 23 psiA |
Test Chamber Pressure Specification: | ±0.0023 psiA |
Display Resolution: | 1E-3 |
Test Chamber Dimensions: | 12" H x 12" W x 12" D (30.5 cm x 30.5 cm x 30.5 cm) |
Physical Dimensions: | 38.25” H x 60” W x 36” D (97.15 cm x 152.4 cm x 91.44 cm) |