Comparison Table

Thermometers Comparison Table
UT-ONE L01AUT-ONE S01AUT-ONE B03BUT-ONE S04AUT-ONE S04BUT-ONE S12AUT-ONE S12BUT-ONE S16B
Number of main channels11344121216
Platinum resistance thermometersselectable*yesyesyesyesyesyesyes
Thermistorsselectable**yesyesyesyesyesyesyes
Thermocouplesnoyesyesyesyesyesyesyes
4-wire resistance measurementyesyesyesyesyesyesyesyes
3-wire resistance measurementnonoyesnoyesnoyesyes
2-wire resistance measurementnonoyesnoyesnoyesyes
Keep-warm currentsnot applicablenot applicableautomaticautomaticautomaticautomaticautomaticautomatic
PRT range0 to 1000 Ω*0 to 1000 Ω0 to 1000 Ω0 to 1000 Ω0 to 1000 Ω0 to 1000 Ω0 to 1000 Ω0 to 1000 Ω
PRT Measurement currents1 mA, 0.707 mA*1 mA, 0.707 mA1 mA, 0.707 mA1 mA, 0.707 mA1 mA, 0.707 mA1 mA, 0.707 mA1 mA, 0.707 mA1 mA, 0.707 mA
Thermistor Range0 to 100 kΩ**0 to 50 kΩ0 to 100 kΩ0 to 50 kΩ0 to 100 kΩ0 to 50 kΩ0 to 100 kΩ0 to 100 kΩ
Thermistor measurement currents10 μA, 7.07 μA**20 μA, 14.1 μA10 μA, 7.07 μA20 μA, 14.1 μA10 μA, 7.07 μA20 μA, 14.1 μA10 μA, 7.07 μA10 μA, 7.07 μA
Built-in displaynonoyesnonononono
LabVIEW driversyesyesyesyesyesyesyesyes
UT-ONE Applications softwareyesyesyesyesyesyesyesyes
Excel Acquisition softwareyesyesyesyesyesyesyesyes
MODBUS RS232 protocolyesnoyesoptionaloptionaloptionaloptionaloptional
SD Card internal memorynonoyesnonononono
Flash internal memoryyesnodepends***noyesnoyesyes
Galvanic isolationnonodepends***noyesnoyesyes
Internal Li-ion batterynonoyesnonononono
USB interfacenoyesyesyesyesyesyesyes
RS232 interfaceyesnoyesoptionaloptionaloptionaloptionaloptional
GPIB+RS232 interfacenononooptionaloptionaloptionaloptionaloptional
Ambient conditions probe connectornonoyesoptionalyesoptionalyesyes
Ambient conditions probenonooptionaloptionaloptionaloptionaloptionaloptional
PRT banana connectornoyesyesyesyesyesyesyes
Thermocouple banana connectornoyesyesyesyesyesyesyes
Miniature thermocouple connectornoyesyesnonononono
IP68 circular connectoryesnonononononono
Auxiliary thermocouple inputnonoyesnoyesnoyesyes
Extended operating temperature rangeyesnonononononono
Waterproofyesnonononononono
PRT accuracy (typical)0.004 °C*0.006 °C0.004 °C0.006 °C0.004 °C0.006 °C0.004 °C0.004 °C
PRT effective resolution0.0005 °C*0.0005 °C0.0005 °C0.0005 °C0.0005 °C0.0005 °C0.0005 °C0.0005 °C
Thermistor accuracy (typical)0.001 °C**0.001 °C0.001 °C0.001 °C0.001 °C0.001 °C0.001 °C0.001 °C
Thermistor effective resolution0.0001 °C**0.0001 °C0.0001 °C0.0001 °C0.0001 °C0.0001 °C0.0001 °C0.0001 °C
Thermocouple accuracy (typical)not aplicable0.1 °C0.1 °C0.1 °C0.1 °C0.1 °C0.1 °C0.1 °C
Thermocouple effective resolutionnot aplicable0.01 °C0.01 °C0.01 °C0.01 °C0.01 °C0.01 °C0.01 °C

* only available on UT-ONE L01A with Pt100 option
** only available on UT-ONE L01A with Thermistor option
*** available as standard feature on UT-ONE B03B units manufactured in 2022 or later (with “Isolated input option” marking on product label)